Skip to content

Pipeline sensitivity and precision

Variant Type Measure Sensitivity Precision Truth set
Single Nucleotide Variants Mean 0.999 0.996 NA12878 with Genome in a Bottle truth set (high confidence regions)
Single Nucleotide Variants 95% credible interval 0.999-0.999 0.996-0.996 NA12878 with Genome in a Bottle truth set (high confidence regions)
Indels Mean 0.936 0.949 NA12878 with Genome in a Bottle truth set (high confidence regions)
Indels 95% credible interval 0.935-0.936 0.948-0.950 NA12878 with Genome in a Bottle truth set (high confidence regions)
Copy Number Variants (>2Kb) Mean 0.975 N/A Clinically significant CNVs detected by standard of care tests in accredited labratories
Copy Number Variants (>2Kb) 95% credible interval 0.956-0.991 N/A Clinically significant CNVs detected by standard of care tests in accredited laboratories
Short Tandem Repeat Expansions Mean 0.959 0.986 STR expansion tests (positive and negative) at target loci performed in accredited laboratories
Short Tandem Repeat Expansions 95% credible interval 0.896-1 0.947-1 STR expansion tests (positive and negative) at target loci performed in accredited laboratories